BLB3 – In-line and Continuous Needle Purge Type

Features

  • Sampling from low pressure devices or process lines: 0-145 psig (0-10 bar)
  • In-line sampling
  • Sampling for viscous liquids
  • Needle purge

Basic Configuration

Wetted Material316 SS
Sleeve Assembly 250 ml sleeve with bottle retaining clip
Needle Assembly Process/vent needle ID: 1.4 mm (0.06″)
Sampling ValveIn-line needle valve:
PTFE packing and PCTFE seat
Max. working pressure: 276 psig @ 70℉ (19 bar @ 20℃)
Temperature range: -18℉ to 298℉ (-28℃ to 148℃)
Nitrogen BranchNitrogen regulator
CV Series check valves, NB Series needle valves
Pressure gauge
ConnectionsProcess: NPS 3/4, ANSI B16.5 Class 150 RF flange
Vent/purge: 1/4″ tube fitting
OthersSpring return handle, purge connection

Typical Installation Mod

1- Preparation

Place a new septum on the sample bottle. Insert the bottle with cap and septum into the sleeve until the septum is pierced by the needles. Swing down the bottle retaining clip.

2- Sampling

open the in-line needle valve, allowing the sample to flow into the bottle. When the required amount has been taken, release the handle to close the valve automatically.

3 – Needle Purge

Open the valve on the Nitrogen branch, allowing Nitrogen to force the residual sample from the system into the bottle.

4 – Off

Close the valve on the Nitrogen branch. Remove the bottle retaining clip and take out the bottle from the sleeve. The septum reseals automatically to complete the sampling process.