Sampling System – BLB3
BLB3 – In-line and Continuous Needle Purge Type
Features
- Sampling from low pressure devices or process lines: 0-145 psig (0-10 bar)
- In-line sampling
- Sampling for viscous liquids
- Needle purge
Basic Configuration
| Wetted Material | 316 SS |
| Sleeve Assembly | 250 ml sleeve with bottle retaining clip |
| Needle Assembly | Process/vent needle ID: 1.4 mm (0.06″) |
| Sampling Valve | In-line needle valve: PTFE packing and PCTFE seat Max. working pressure: 276 psig @ 70℉ (19 bar @ 20℃) Temperature range: -18℉ to 298℉ (-28℃ to 148℃) |
| Nitrogen Branch | Nitrogen regulator CV Series check valves, NB Series needle valves Pressure gauge |
| Connections | Process: NPS 3/4, ANSI B16.5 Class 150 RF flange Vent/purge: 1/4″ tube fitting |
| Others | Spring return handle, purge connection |
Typical Installation Mod

1- Preparation
Place a new septum on the sample bottle. Insert the bottle with cap and septum into the sleeve until the septum is pierced by the needles. Swing down the bottle retaining clip.
2- Sampling
open the in-line needle valve, allowing the sample to flow into the bottle. When the required amount has been taken, release the handle to close the valve automatically.
3 – Needle Purge
Open the valve on the Nitrogen branch, allowing Nitrogen to force the residual sample from the system into the bottle.
4 – Off
Close the valve on the Nitrogen branch. Remove the bottle retaining clip and take out the bottle from the sleeve. The septum reseals automatically to complete the sampling process.