BLA2 – System Purge Type

Features

  • Sampling from low pressure devices or process lines:0-145 psig (0-10 bar)
  • System purge

Basic Configuration

Wetted Material 316 SS
Sleeve Assembly 250 ml sleeve with bottle retaining clip
Needle Assembly Process/vent needle ID: 1.4 mm (0.06″)
Sampling Valve BF Series 3-way ball valves:
PTFE seat and FKM O-ring
Max working pressure: 1500 psig @ 70℉ (103 bar @ 20℃)
Temperature range: 0℉ to 450℉ (-18℃ to 232℃)
Connections1/4″ tube fitting

Typical Installation Mode

1- Preparation

Place a new septum on the sample bottle. Insert the bottle with cap and septum into the sleeve until the septum is pierced by the needles. Swing down the bottle retaining clip.

2 – System Purge

Turn the handle to the “BYPASS” position, allowing the sample to flow continuously through the system and purge system to ensure representative sampling.

3 – Sampling

Turn the handle to “SAMPLE” position, allowing the sample to flow into the bottle. When the required amount has been taken, turn the handle to the “OFF” position to close the sampling system.

4- Off

Remove the bottle retaining clip and take out the bottle from the sleeve. The septum reseals automatically to complete the sampling process.